SA-6 Class 1 Medical Device: Medical Gas Pipeline Systems.
SA-6 Computer Interface Manual
For more information & documentation please contact us
SA-6 Area Alarm
Medical Gas Pipeline Area Alarm System
Features
The SA-6 Area Alarm is a simple, robust and proven medical gas monitoring system designed for long-term reliability and ease of installation.
Since its introduction over 30 years ago, the SA-6 Area Alarm has been re-issued on a number of occasions. Each revision has maintained PCB compatibility with previous versions, ensuring continuity for installed systems.
With over 30,000 units installed worldwide, the SA-6 is widely regarded as the industry standard for ward-level gas monitoring. Shire Controls continues to manufacture and support the SA-6 with no plans to de-list the product.
Application
The SA-6 Area Alarm is used to monitor medical gas pipeline pressure within wards, theatres and clinical areas.
Pressure switches are installed downstream of the final AVSU and monitor:
The system continuously monitors the integrity of the cabling between the pressure switches and the alarm panel. A cable fault will result in:
Connection
The simple analogue design allows inputs to be connected via end-of-line termination units or directly from contact sources.
A multi-core cable is run from the end-of-line transmitter (or pressure switch) to the alarm panel, comprising:
The SA-6 can be configured to operate:
Please specify configuration requirements at time of order.
Termination Boards
To enable cable monitoring, a termination board must be installed as close as practical to the pressure switches.
Termination boards are supplied as:
When the SA-6 is installed within a zone module or AVSU enclosure where the pressure switches are located in the same enclosure as the alarm panel, cable monitoring is not required, and termination boards are therefore not necessary.
Master–Repeater Configuration
The SA-6 supports a master–repeater configuration:
If two repeaters are used, the wiring must be:
Master → Repeater 1 → Repeater 2
Installation Requirements
Installation must be carried out by competent personnel in accordance with HTM 02-01 and local regulations.
Electrical Supply
Signal Cabling
Optional Equipment
Equipment Warranty
Standard Warranty
Alarms returned to our works which have failed due to faulty parts or manufacture will be repaired or replaced without charge for parts and labour.
10-Year Compatibility Warranty
Install the SA-6 with confidence. Equipment will remain available to modify or extend the system for at least 10 years from installation.
Compliance
HTM 02-01 ✔
ISO 13485 ✔
EN ISO 7396-1 ✔
IEC 62304 (Class A) ✔
UK MDR / EU MDR ✔